ITEM
|
SPEC
|
REMARKS
|
|
Model: MMS3000-T4 Incorporating 4 Temperature Channels
|
|
TEMPERATURE RANGE
|
T Type Thermocouple
|
-418 to +743°F
(-250 to +395°C)
|
Operating temperature range
|
E Type Thermocouple
|
-418 to +1823°F
(-250 to +995°C)
|
is determined by operating
|
J Type Thermocouple
|
-346 to +2183°F
(-210 to +1195°C)
|
range of thermocouple
|
K Type Thermocouple
|
-418 to +2498°F
(-250 to +1370°C)
|
|
TEMPERATURE RESOLUTION
|
|
Above -148°F
(-100°C): 0.1°F
or
°C
|
|
|
At or below -148°F
(-100°C): 1°F
or
°C
|
Effective resolution decreases at or below -148°F
(-100°C)
|
TEMPERATURE ACCURACY
|
|
Above -148°F
(-100°C): ?.1% of reading ?.9°F
(0.5°C)
At or below -148°F (-100°C): ?% of reading ?.9°F
(0.5°C)
|
Total measurement accuracy depends on MMS3000-T4 and thermocouple accuracy. Only MMS3000-T4 accuracy is given here. MMS3000-T4 temperature 73 ?9°F
(23 ?5°C), charger off. See Appendix 3 of Owners Manual
|
MINIMUM SAMPLING INTERVAL
|
|
# of channels selected (User-selectable)
|
Minimum sampling interval (seconds)
|
1
|
0.2
|
2
|
0.3
|
3
|
0.5
|
4
|
0.6
|
|
Sampling interval is programmable from 0.2 seconds to 60 hours
Note: Minimum time to scan all selected channels - not each channel
|
DATA LOGGING
|
Data Storage
|
512 kilobytes ?approx. 400,000 individual samples
|
Non-volatile memory (with battery back-up). # of Samples depends on # of Data Sets and Recordings.
|
Data Storage Format
|
Up to 50 named DATA SETS
|
User-specified name entered from keypad
|
|
Up to 50 RECORDINGS per data set
|
Each recording has unique time/date stamp
|
|
Up to 9,999 SAMPLES per recording
|
Each sample contains data for all selected channels
|
Data Set Name
|
Up to 16 alpha-numeric characters
|
|
DISPLAY
|
|
LCD with graphic capabilities
|
|
Resolution
|
240 x 128 pixels
|
|
Viewing Area
|
4.3" x 2.3" (110mm x 60mm)
|
|
Backlight
|
Electro-luminescent
|
|
PROFLASH
|
|
Allows internal firmware to be upgraded via built-in RS232 port
|
Download firmware upgrades via the Internet
|
COMMUNICATIONS
|
|
RS232
|
15kV ESD Protected. Cable with DB9 connector included
|
Baud Rate
|
9600, 19200, 57600 bits per second
|
Automatic Baud Rate Selection
|
INPUT TERMINATION
|
|
Sockets accept miniature thermocouple twin blade plug
|
Blades spaced 5/16" (7.9mm) center-to-center
|
INPUT IMPEDANCE
|
|
Greater than 100k Ohms
|
Terminal-to-terminal
|
ISOLATION
|
|
50VDC minimum
|
Between temperature inputs and MMS3000-T4 outputs e.g. RS232 port. No isolation between input channels.
|
BATTERY
|
Type
|
Custom Nickel-Cadmium pack
|
|
Voltage
|
7.2 Volts nominal
|
|
Capacity
|
1500 mAh nominal
|
|
Operating Time
|
13 hrs with backlight off, 7 hrs backlight on - nominal
|
|
CHARGER AND CONDITIONER
|
|
Built-in dual rate charger ?Automatic and Manual control
|
AC Adapter included (13.5 ?1.5 VDC, 1A output)
|
Fast Charge Rate
|
0.7A nominal
|
2.5 hours for complete charge nominal
|
Discharge Rate
|
0.5A nominal
|
Combats NiCad battery memory effect
|
MECHANICAL
|
Size
|
9.7"W x 6.1"L x 3.0"H (247mm x 154mm x 75mm)
|
Including protective boot
|
Weight
|
4.4lb (2kg)
|
Including protective boot and strap
|
ENVIRONMENTAL
|
Temperature/Humidity
|
|
|
Operating
|
32°F
to 122°F
(0°C
to 50°C) / 70% RH
|
Non-condensing
|
|
32°F
to 86°F
(0°C
to 30°C) / 80% RH
|
Non-condensing
|
Storage
|
14°F
to 140°F
(-10°C
to 60°C
) / 95% RH
|
|
EMC
|
EN50081-1
|
Radiated and conducted emissions
|
|
EN50082-1
|
ESD Immunity and RF field immunity (EMI).
|
The manufacturer reserves the right to modify specifications without prior notice.
|