Multi-channel, multi-tasking ultrasonic instruments, with advanced networking and reporting capabilities for nondestructive inspection and evaluation
PRODUCT DESCRIPTION

Introducing the 2000 Series
The upi-50 2000 Series extends the capabilities of the industry-standard, fully-digital upi-50 ultrasonic instrument into the realm of high-speed, a multi-tasking applications with full compatibility to networking, data archiving and reporting codes. Encompassing new RPP2 or RPP4 programmable square-wave pulser preamplifiers with up to eight independent channels on board, the upi-50 2000 Series provides the industrial user and development lab with the next-generation platform for ultrasonic applications.
Hardware
Industry-certified UPR-50 digital ultrasonic electronics, one to eight channels, with a 50MHz bandwidth and 100MSamp/sec sampling rate. Choice of Pentium1 or Pentium II1 CPU, with Windows NT2 operating system. Selection of on-board mass-storage devices.
Enclosures
Rack-mount or transportable tabletop enclosure, with full-feature I/O connector panel.
High-resolution screen displays
800 x 600 pixel, extra-bright TFT screen display on tabletop instrument; high-resolution 17” display is standard with rack-mount version.
Front-end square wave pulser
Choice of plug-in RPP2 or RPP4 programmable square-wave pulsers, each with linear preamplifier, offering outstanding near-surface resolution, penetrating power, and equivalent input noise less than 4nvolt/Ö Hz.
Linear or log amplification
Optional plug-in LAM4 programmable logarithmic amplifier with dynamic range exceeding 85dB offers menu-selectable linear or logarithmic signal conditioning with exceptional low-noise operation.
ScanMaster 2000 software
ScanMaster 2000 software for instrument set-up and display, application control, flaw evaluation, measurement and documentation, with the full multi-tasking capability of Windows NT. All instrument functions directly accessible from the ScanMaster 2000 Toolbox with easy-to-learn, easy-to-use application tailored operator interfaces.
Applications
Flaw detection and evaluation in metals and composites, wall and clad thickness measurements, B and C-scan imaging and evaluation.
Report Generator
Documented inspection reports application-tailored to customer specifications.
Visual ScanMaster Builder
Custom-tailored ‘virtual’ control panels for specific inspection applications. Knobs look and work just like analog instrument controls.

 1998 Scan Master 2000 APPLICATIONS
  • Servo control interface for AC and DC servo motors
  • C-scan imaging with encoder synchronization
  • Macro language capabilities - custom tailored building blocks for automated inspection
  • Longitudinal and spiral weld B and C-scan imaging
  • Full-body bar and pipe inspection
  • Transducer characterization with FFT and inspection report
  • Multi-channel wall thickness measurement
Overall Analog Performance
Analog bandwidth,receiver 1 to 50MHz (-3dB)
Calibrated gain 92dB
Gain step size 0.2dB
Full screen sensitivity 50m V peak to peak
Equivalent input noise < 4nV Hz
Linearity ± 1dB
Channels
Number of channels 1 to 8
Channel triggering Internal or external
Front End RPP2 and RPP4 Pulsers
Installation on-board 1 to 4 ch. 1) - RPP2
  1 to 8 ch. - RPP4
Free-standing or mounted in Octaport 1 to 8 ch., installation to 50 m (150 ft)
Pulser type High resolution (HR)

High voltage (HV)
RPP2 Pulser Characteristics
Excitation Square wave
Amplitude 8 levels
Max. pulse amplitude, 50W load 300V + 10% (HR) 450V + 10% (HV)
PRF 2 to 5000Hz
Damping 8 settings
Stability ± 2% FSH
Flatness ± 0.5dB
Mode PE, TT, Fire Twice, Fire Next
Pulse rise time 9nsec (RPP2 HR)
  25nsec (RPP2 HV)
Pulse width 16-500nsec (HR)
  50-500nsec (HV)
No. of steps 250
Isolation (PE/TT) > 60dB @ 5MHz
Preamplifier Characteristics
Programmable gain 0, 12, 25, 42dB
Bandwidth (-3dB) 1-35MHz
Output voltage 2.0Vp-p

LAM4 Logarithmic Amplifier
Installation In series with RPP2
Dynamic range 85 dB
Bandwidth 1-10MHz
Output Voltage 1.0Vp-p
Frequency Filters
Number and type of filters 7 band pass, 1 wide-band
Time Base
Range 2) 3) 100nsec to 245m sec
Delay plus range2) 3) 100nsec to 640m sec
Resolution 2) 10nsec
Hardware Gates
Number 4, expandable to 32
Modes Master or slave
Range 2) 3) 100nsec to 245m sec
Delay and range 2) 3) 120nsec to 245m sec
Resolution 3) 10nsec
Peak detection Positive, negative or absolute
Alarm threshold Positive or negative going
Dynamic gate Backwall echo tracking
DAC
Dynamic range 50dB
Amplitude resolution 0.2dB
Number of steps 1 to 506
Step width 3) 320/640nsec
Position resolution 3) 10nsec (first step)
Slew rate > 40dB/320nsec
Interpolation Between calibration points
Last gate attenuator To 50dB
Analog to Digital Conversion
Resolution 8 bit
Sampling rates 6.25,12.5,25,50,100Msamp/sec, single shot
High speed buffer 24 or 48KByte

Data acquisition speed
Peak amplitude and time of flight Up to 132K measurements/sec
A-scan capture Up to 10MByte/sec continuous
Computer, Data Storage and Display
CPU Pentium or Pentium II, from 233MHz
RAM 32MByte, expansion to 128MByte
Hard disc drive 2.1GByte, expansion to 6.1GByte
Floppy disk drive 1.44MByte
Display 10.4” high-brightness TFT color (800 x 600 pixels)
Control device Keyboard, mouse or trackball
Remote control Programmable hand-held infra-red
Removable mass storage device 100MByte, 1GByte
Interfaces  
Standard SCSI
  External VGA
  Printer, Dual serial
Optional 2 or 16 port digital I/0
  Dual encoder
  LAN, MODEM
  2 or 8 Ch. D/A
  2 or 8 Ch. A/D
General
Power requirements 115V or 230V ± 10%
  48-63Hz, 250VA
Operating temperature range 5 to 40° C - RPP2,4 4)

5 to 35° C - upi-50
Humidity < 80% non-condensing
Dimensions 450mm x 220mm x 390mm (17.7” x 8.7” x 15.4”) - W:H:D
Weight, single channel version 13.5Kg (30Lbs)

1) 1 - 6 ch. in Fire Twice mode.   2) With standard 24K buffer. Maximum range and delay doubled with 48K buffer.   3) At 100MSamp/sec sampling frequency.   4) Mounted externally.